Bringing a repeatable temperature source to the plant floor, this dry well calibrator spans 33 to 660°C, displays within ±0.3°C to 420°C and ±0.5°C at 660°C, and reaches 660°C in 15 minutes. Insert H sets the hole pattern in the well. Auto step, ramp and soak run from the 6.5 in touch screen, with Wi-Fi and Ethernet for remote control.
Bringing a repeatable temperature source to the plant floor, this dry well calibrator spans 33 to 660°C, displays within ±0.3°C to 420°C and ±0.5°C at 660°C, and reaches 660°C in 15 minutes. Insert H sets the hole pattern in the well. Auto step, ramp and soak run from the 6.5 in touch screen, with Wi-Fi and Ethernet for remote control.
Features
Applications
| Temperature Performance | |
| Temperature Range at 23°C | 33 to 660°C |
| Display Accuracy | ±0.3°C at 33°C, ±0.3°C at 420°C, ±0.5°C at 660°C |
| Stability | ±0.02°C at 33°C, ±0.03°C at 50°C, ±0.04°C at 420°C, ±0.04°C at 660°C |
| Axial Uniformity at 60 mm (2.4") | ±0.05°C at 33°C, ±0.3°C at 420°C, ±0.5°C at 660°C |
| Radial Uniformity | ±0.02°C at 33°C, ±0.05°C at 420°C, ±0.1°C at 660°C |
| Loading Effect | ±0.15°C display sensor, ±0.035°C external sensor |
| Hysteresis | 0.1°C (display sensor) |
| Resolution | 0.01°C |
| Temperature Units | °C, °F and K |
| Heating Time | 15 minutes from 33 to 660°C |
| Cooling Time | 23 minutes from 660 to 100°C, 12 minutes from 100 to 50°C, 12 minutes from 50 to 33°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Immersion Depth | 5.90" (150 mm) |
| Insert Outside Diameter | 0.98" (24.8 mm) |
| Insert Supplied | ADT110-875-H-INSERT-H |
| General | |
| Display | 6.5" (165 mm) color touch screen |
| Communication | USB-A, USB-B, RJ45, Wi-Fi, Bluetooth |
| Dimensions | 12.6 x 6.7 x 13.0" (320 x 170 x 330 mm) |
| Weight | 18.9 lb (8.6 kg) |
| Power Requirements | 90 to 254 V AC, 45 to 65 Hz, 1200 W |
| Environmental Conditions | 8 to 38°C for guaranteed accuracy, 0 to 50°C operating, 0% to 90% RH non-condensing, to 3000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g (10 to 500 Hz) for 30 minutes on 2 sides; impact 4 g three times; drop test 19.6" (500 mm) |
| Temperature Performance | |
| Temperature Range at 23°C | 33 to 660°C |
| Display Accuracy | ±0.3°C at 33°C, ±0.3°C at 420°C, ±0.5°C at 660°C |
| Stability | ±0.02°C at 33°C, ±0.03°C at 50°C, ±0.04°C at 420°C, ±0.04°C at 660°C |
| Axial Uniformity at 60 mm (2.4") | ±0.05°C at 33°C, ±0.3°C at 420°C, ±0.5°C at 660°C |
| Radial Uniformity | ±0.02°C at 33°C, ±0.05°C at 420°C, ±0.1°C at 660°C |
| Loading Effect | ±0.15°C display sensor, ±0.035°C external sensor |
| Hysteresis | 0.1°C (display sensor) |
| Resolution | 0.01°C |
| Temperature Units | °C, °F and K |
| Heating Time | 15 minutes from 33 to 660°C |
| Cooling Time | 23 minutes from 660 to 100°C, 12 minutes from 100 to 50°C, 12 minutes from 50 to 33°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Immersion Depth | 5.90" (150 mm) |
| Insert Outside Diameter | 0.98" (24.8 mm) |
| Insert Supplied | ADT110-875-H-INSERT-H |
| General | |
| Display | 6.5" (165 mm) color touch screen |
| Communication | USB-A, USB-B, RJ45, Wi-Fi, Bluetooth |
| Dimensions | 12.6 x 6.7 x 13.0" (320 x 170 x 330 mm) |
| Weight | 18.9 lb (8.6 kg) |
| Power Requirements | 90 to 254 V AC, 45 to 65 Hz, 1200 W |
| Environmental Conditions | 8 to 38°C for guaranteed accuracy, 0 to 50°C operating, 0% to 90% RH non-condensing, to 3000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g (10 to 500 Hz) for 30 minutes on 2 sides; impact 4 g three times; drop test 19.6" (500 mm) |