Made for accurate gas flow measurement, this gas mass flow meter reads 0.25 to 50 SLM with ±0.8% accuracy. It pairs a 0 to 5 V analog output and RS-232 communication with an on-board LCD readout. Rated to 145 psi (10 bar) through a 1/4" NPT connection, it works great for process and laboratory gas monitoring.
Made for accurate gas flow measurement, this gas mass flow meter reads 0.25 to 50 SLM with ±0.8% accuracy. It pairs a 0 to 5 V analog output and RS-232 communication with an on-board LCD readout. Rated to 145 psi (10 bar) through a 1/4" NPT connection, it works great for process and laboratory gas monitoring.
The FMA-1600A Series mass and volumetric flow meters use the principle of differential pressure within a laminar flow field to determine the mass flow rate. A laminar flow element (LFE) inside the meter forces the gas into laminar (streamlined) flow. Inside this region, the Poiseuille equation dictates that the volumetric flow rate be linearly related to the pressure drop.
A differential pressure sensor is used to measure the pressure drop along a fixed distance of the LFE. This, along with the viscosity of the gas, is used to accurately determine the volumetric flow rate. Separate absolute temperature and pressure sensors are incorporated and correct the volumetric flow rate to a set of standard conditions.
This standardized flow rate is commonly called the mass flow rate and is reported in units such as standard cubic feet per minute (SCFM) or standard liters per minute (SLM).
The gas select feature can be adjusted from the front keypad or via RS232 communications. Volumetric flow, mass flow, absolute pressure, and temperature can all be viewed or recorded through the RS232 connection.
It is also possible to multi-drop up to 26 units on the same serial connection to a distance of 125' (38 m). These flow meters are available in a portable version ("-B" option), the battery charge will last up to 18 hours.
Features
Applications
| Accuracy | ±(0.8% of rdg + 0.2% FS) |
| Repeatability | ±0.2% |
| Turndown ratio | 200:1 |
| Response time | 10 ms typical default response time for 63.2% of a step change, field adjustable via RS232 communications |
| Output | 0 to 5 Vdc standard |
| Operating temperature | 14 to 122°F (-10 to 50°C) |
| Zero shift | 0.02% FS/°C/atm |
| Span shift | 0.02% FS/°C/atm |
| Humidity range | 0 to 100% non-condensing |
| Pressure (maximum) | 145 psig |
| Measurable flow rate | 125% FS |
| Supply voltage | 7 to 30 Vdc (15 to 30 Vdc for 4 to 20 mA output) |
| Supply current | 35 mA typical current draw; 100 mA available supply recommended |
| Cable connection | 8-pin mini DIN |
| Wetted parts | 302 and 303 SS, FKM, heat cured silicone RTV, glass reinforced PPS, heat cured epoxy, aluminum, gold, brass, 430 FR stainless (416 stainless steel on large sizes), silicon, glass |
| Connection | 1/4 FNPT |
| Maximum flow | 50 SLM |
| Dimensions | 4.367" (110.92 mm) Length: 4.0" (101.60 mm) Depth: 1.6" (40.64 mm) Port size: 1/4 FNPT |
| Accuracy | ±(0.8% of rdg + 0.2% FS) |
| Repeatability | ±0.2% |
| Turndown ratio | 200:1 |
| Response time | 10 ms typical default response time for 63.2% of a step change, field adjustable via RS232 communications |
| Output | 0 to 5 Vdc standard |
| Operating temperature | 14 to 122°F (-10 to 50°C) |
| Zero shift | 0.02% FS/°C/atm |
| Span shift | 0.02% FS/°C/atm |
| Humidity range | 0 to 100% non-condensing |
| Pressure (maximum) | 145 psig |
| Measurable flow rate | 125% FS |
| Supply voltage | 7 to 30 Vdc (15 to 30 Vdc for 4 to 20 mA output) |
| Supply current | 35 mA typical current draw; 100 mA available supply recommended |
| Cable connection | 8-pin mini DIN |
| Wetted parts | 302 and 303 SS, FKM, heat cured silicone RTV, glass reinforced PPS, heat cured epoxy, aluminum, gold, brass, 430 FR stainless (416 stainless steel on large sizes), silicon, glass |
| Connection | 1/4 FNPT |
| Maximum flow | 50 SLM |
| Dimensions | 4.367" (110.92 mm) Length: 4.0" (101.60 mm) Depth: 1.6" (40.64 mm) Port size: 1/4 FNPT |