Set of 4x 20/200A switchable mini current clamps for Fluke 1735.
Set of 4x 20/200A switchable mini current clamps for Fluke 1735.
| Nominal Current | 20 / 200 A ac rms |
| Overload | Up to 260 A ac rms |
| Maximum Non-Destructive Current | 300 A ac rms |
| Conductor Position Influence | < 0.5% of range at 50/60 Hz |
| Error Due to Adjacent Conductor | ≤ 15 mA/A at 50 Hz |
| Phase Shift at 2 kHz at 200 A | < ±1 degree |
| Frequency Range | 40 Hz to 5 kHz (-3 dB) |
| Temperature Coefficient | 0.015 % of range/°C |
| Measuring range | 5 to 95 %RH |
| Working Voltage | ±3 %RH |
| Accuracy at 50Hz | 0.5 to 10A: ±2.5% 10 to 50A: ±1.5%, <2° phase shift 50 to 100A: ±0.5%, <0.5° phase shift 100 to 200A: ±0.5%, <1° phase shift |
| Maximum Conductor Size | 15 mm diameter, bus bar 15 mm x 17mm |
| Cable Length | 2m |
| Safety Standards | 600 V CAT III, pollution degree 2 IEC/EN61010-1: 2001 IEC/EN61010-2-032:2002 IEC/EN61010-031:2002 |
| EMC Standards | IEC/EN 61326-1:2006 |
Click on a category to view a selection of compatible accessories with the Fluke I20/200A-CLAMP-PQ4 4-phase 20A/200A Mini Clamp-On Current Probes.