This Inconel-sheathed secondary reference temperature standard features a durable design thanks to its fully supported sensing element, which is surrounded in a ceramic case with a moisture- and contaminant-resistant hermetic glass seal. The temperature standard is ideal for use in the factory, field or lab and features spade lug connections and 9 x 0.188in sizing.
This Inconel-sheathed secondary reference temperature standard features a durable design thanks to its fully supported sensing element, which is surrounded in a ceramic case with a moisture- and contaminant-resistant hermetic glass seal. The temperature standard is ideal for use in the factory, field or lab and features spade lug connections and 9 x 0.188in sizing.
It bridges the gap between a 100Ω industrial RTD and an SPRT.
Features
| Temperature Range | -328 to 788°F (-200 to 420°C) | 
| Nominal Resistance At 0°C | 100Ω ±0.10Ω | 
| Temperature Coefficient | 0.0039250Ω/Ω/°C | 
| Accuracy (Includes Calibration And 100-Hour Drift (k=2)) | ±0.024°C at -200°C; ±0.012°C at 0°C; ±0.035°C at 420°C | 
| Drift (After 100 Hours At Max Temp, 95% Confidence (k=2)) | ±0.007°C at 0.010°C | 
| Sensor Length | 1.1in (28mm) | 
| Sensor Location | 0.27in ±0.13in (6.9mm ±3.3 mm) from tip | 
| Lead-Wire Cable Type | PTFE insulated with PTFE jacket, 22AWG | 
| Lead-Wire Length | 6ft (183cm) | 
| Calibration | NVLAP-accredited calibration included, lab code 200706-0 | 
Click on a category to view a selection of compatible accessories with the Fluke 5615-9-S Secondary Reference Temperature Standards with Spade Lugs, 9 x 0.188in.
| Temperature Range | -328 to 788°F (-200 to 420°C) | 
| Nominal Resistance At 0°C | 100Ω ±0.10Ω | 
| Temperature Coefficient | 0.0039250Ω/Ω/°C | 
| Accuracy (Includes Calibration And 100-Hour Drift (k=2)) | ±0.024°C at -200°C; ±0.012°C at 0°C; ±0.035°C at 420°C | 
| Drift (After 100 Hours At Max Temp, 95% Confidence (k=2)) | ±0.007°C at 0.010°C | 
| Sensor Length | 1.1in (28mm) | 
| Sensor Location | 0.27in ±0.13in (6.9mm ±3.3 mm) from tip | 
| Lead-Wire Cable Type | PTFE insulated with PTFE jacket, 22AWG | 
| Lead-Wire Length | 6ft (183cm) | 
| Calibration | NVLAP-accredited calibration included, lab code 200706-0 | 
Click on a category to view a selection of compatible accessories with the Fluke 5615-9-S Secondary Reference Temperature Standards with Spade Lugs, 9 x 0.188in.