Holding laboratory-grade stability and uniformity in a portable well, this reference dry well calibrator spans -40 to 160°C, holds ±0.1°C across the full range, and reaches 160°C in 10 minutes. Insert E sets the hole pattern in the well. Auto step, ramp and soak run from the 6.5 in touch screen, with Wi-Fi and Ethernet for remote control.
Holding laboratory-grade stability and uniformity in a portable well, this reference dry well calibrator spans -40 to 160°C, holds ±0.1°C across the full range, and reaches 160°C in 10 minutes. Insert E sets the hole pattern in the well. Auto step, ramp and soak run from the 6.5 in touch screen, with Wi-Fi and Ethernet for remote control.
Features
Applications
| Temperature Performance | |
| Temperature Range at 23°C | -40 to 160°C |
| Display Accuracy | ±0.1°C at full range |
| Stability | ±0.005°C at full range |
| Axial Uniformity at 60 mm (2.4") | ±0.035°C at -40°C, ±0.020°C at 0°C, ±0.050°C at 160°C |
| Axial Uniformity at 80 mm (3.15") | ±0.050°C at -40°C, ±0.040°C at 0°C, ±0.050°C at 160°C |
| Radial Uniformity | ±0.01°C at full range |
| Loading Effect | ±0.08°C display sensor, ±0.010°C external sensor |
| Hysteresis | 0.025°C (display sensor) |
| Resolution | 0.001°C |
| Temperature Units | °C, °F and K |
| Heating Time | 4 minutes from -40 to 23°C, 10 minutes from 23 to 160°C |
| Cooling Time | 8 minutes from 160 to 23°C, 15 minutes from 23 to -40°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Immersion Depth | 6.30" (160 mm) |
| Insert Outside Diameter | 1.26" (31.9 mm) |
| Insert Supplied | ADT110-878-L-INSERT-E |
| General | |
| Display | 6.5" (165 mm) color touch screen |
| Communication | USB-A, USB-B, RJ45, Wi-Fi, Bluetooth |
| Dimensions | 6.69 x 13.58 x 13.0" (170 x 345 x 330 mm) |
| Weight | 24.7 lb (11.2 kg) |
| Power Requirements | 90 to 254 V AC, 45 to 65 Hz, 580 W |
| Environmental Conditions | 8 to 38°C for guaranteed accuracy, 0 to 50°C operating, 0% to 90% RH non-condensing, to 3000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g (10 to 500 Hz) for 30 minutes on 2 sides; impact 4 g three times; drop test 19.6" (500 mm) |
| Temperature Performance | |
| Temperature Range at 23°C | -40 to 160°C |
| Display Accuracy | ±0.1°C at full range |
| Stability | ±0.005°C at full range |
| Axial Uniformity at 60 mm (2.4") | ±0.035°C at -40°C, ±0.020°C at 0°C, ±0.050°C at 160°C |
| Axial Uniformity at 80 mm (3.15") | ±0.050°C at -40°C, ±0.040°C at 0°C, ±0.050°C at 160°C |
| Radial Uniformity | ±0.01°C at full range |
| Loading Effect | ±0.08°C display sensor, ±0.010°C external sensor |
| Hysteresis | 0.025°C (display sensor) |
| Resolution | 0.001°C |
| Temperature Units | °C, °F and K |
| Heating Time | 4 minutes from -40 to 23°C, 10 minutes from 23 to 160°C |
| Cooling Time | 8 minutes from 160 to 23°C, 15 minutes from 23 to -40°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Immersion Depth | 6.30" (160 mm) |
| Insert Outside Diameter | 1.26" (31.9 mm) |
| Insert Supplied | ADT110-878-L-INSERT-E |
| General | |
| Display | 6.5" (165 mm) color touch screen |
| Communication | USB-A, USB-B, RJ45, Wi-Fi, Bluetooth |
| Dimensions | 6.69 x 13.58 x 13.0" (170 x 345 x 330 mm) |
| Weight | 24.7 lb (11.2 kg) |
| Power Requirements | 90 to 254 V AC, 45 to 65 Hz, 580 W |
| Environmental Conditions | 8 to 38°C for guaranteed accuracy, 0 to 50°C operating, 0% to 90% RH non-condensing, to 3000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g (10 to 500 Hz) for 30 minutes on 2 sides; impact 4 g three times; drop test 19.6" (500 mm) |